TECHNICAL DATA
64 MAX, 62 MAX+ and 62 MAX
IR Thermometers
Takes the heat, the dust and water and a 3 meter
drop. And now, internal memory and unattended
monitoring.
TEMPERATURE RANGE
64 MAX
-30 °C to 600 °C (-22 °F to 1112 °F)
62 MAX+
-30 °C to 650 °C (-22 °F to 1202 °F)
62 MAX
-30 °C to 500 °C (-22 °F to 932 °F)
TEMPERATURE ACCURACY
64 MAX
± 1.0 °C or ± 1.0 % of reading,
whichever is greater
± 2.0 °C at -10 °C to 0 °C
± 3.0 °C at -30 °C to -10 °C
62 MAX+
± 1.0 °C or ± 1.0 % of reading,
whichever is greater
± 2.0 °C at -10 °C to 0 °C
± 3.0 °C at -30 °C to -10 °C
62 MAX
± 1.5 °C or ± 1.5 % of reading,
whichever is greater
± 2.0 °C at -10 °C to 0 °C
± 3.0 °C at -30 °C to -10 °C
The Fluke 64 MAX, 62 MAX+ and 62 MAX IR Thermometers have the
precision you need to do your job accurately and will not break the
budget. Designed and tested to survive a 3 meter drop, you can count
on these lightweight, compact infrared thermometers to work when
you need them to work–even in the harshest environments and when
you’re not able to be there (64 MAX).
• Precise laser technology for more accurate and repeatable
measurements
• Temperature accuracy of up to ≥ 0 °C: ± 1 °C or ± 1 % of reading
whichever is greater with 20:1 distance to spot ratio (64 MAX)
• Flashlight (64 MAX) and large, easy-to-read backlit LCD display for
easy viewing even in dark environment
• IP54 rated for extra protection against airborne contaminants
• Get your tool to work when you can’t–set time and desired interval
between measurements and Auto Capture will capture spot temperatures unattended (64 MAX)
• 99 data point logging (64 MAX)
• Displays the minimum, maximum or average temperature, or the
difference between two measurements
DISTANCE:SPOT
64 MAX
20:1
• Hi and Lo alarms for rapid display of measurements outside set
limits
62 MAX+
12:1
• The 62 MAX+ has dual rotating lasers to help you identify area to
be measured. The measurement area is the spot between the dots
(62 MAX and 64 MAX have 1).
62 MAX
10:1
INTERNAL MEMORY
64 MAX
99 data point logging
AUTO CAPTURE
64 MAX
[email protected]
• Small and lightweight, fits easily into your tool box
Fluke-Direct .com
1.888.475.5235
Specifications
64 MAX
62 MAX+
62 MAX
- 30 °C to 600 °C
(-22 °F to 1112 °F)
- 30 °C to 650 °C
(-22 °F to 1202 °F)
- 30 °C to 500 °C
(-22 °F to 932 °F)
Key features
Temperature range
Accuracy (Calibration geometry with ambient
temperature 23 °C +/- 2 °C)
± 1.5 °C or ± 1.5 % of
reading, whichever is
greater
± 2.0 °C at -10 °C to 0 °C
± 3.0 °C at -30 °C to
-10 °C
± 1.0 °C or ± 1.0 % of reading, whichever is greater
± 2.0 °C at -10 °C to 0 °C
± 3.0 °C at -30 °C to -10 °C
Optical resolution
20:1
(calculated at 90 %
energy)
12:1
(calculated at 90 %
energy)
10:1
(calculated at 90 %